Model Number FV251T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: no significant deformations or damage of the valve was detected during the visual inspection.Permeability test: a permeability test has shown that the shuntassitant is permeable.Computer controlled test: to investigate the drainage, the out-flow is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve operates within the accepted tolerance in the vertical positions.Internal inspection: after dismantling of the valve, deposits were found in shuntassistant.To make the proteins / deposits in the shuntassistant more visible, they were colored using a staining solution.Results: based on our investigation results, we can determine deposits.The deposits had no affect to the technical properties at the time of the investigation.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a shuntassistant (part # fv251t) was implanted during a procedure performed on an unknown date.According to the complainant, the valve was believed to be difficult to adjust.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 6 years, height: 118 centimeters (cm), weight: 27 kilograms (kg), gender: female.Same event as mdr id # 3004721439-2022-00365.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNTASSISTANT 20
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15936661
Report Number3004721439-2022-00366
Device Sequence Number12317059
Product Code JXG
UDI-Device Identifier04041906113209
UDI-Public4041906113209
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2022
Device Model NumberFV251T
Device Catalogue NumberFV251T
Device Lot Number20035932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Initial Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexFemale
Patient Weight27 KG
-
-