| Model Number |
324906 |
| Medical Device Problem Code |
Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that 5 bd veo¿ insulin syringes with bd ultra-fine¿ needles experienced product damage while still considered operable.The following information was provided by the initial reporter: consumer reported found the top of syringe to 5 units were melted.Caller drew into syringe humalog and then noticed the outside of barrel to be melted.Caller did not inject with this syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 5 bd veo¿ insulin syringes with bd ultra-fine¿ needles experienced product damage while still considered operable.The following information was provided by the initial reporter: consumer reported found the top of syringe to 5 units were melted.Caller drew into syringe humalog and then noticed the outside of barrel to be melted.Caller did not inject with this syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for evalution?yes,.D10: returned to manufacturer on: 03-jan-2023 h6: investigation summary customer returned two syringes 0.3ml 6mm 90 bx 450 mo.Customer reported that they found the top of syringe to the 5 unit to be melted.Caller drew into syringe humalog and then noticed the outside of barrel to be melted.Caller did not inject with this syringe.However, the syringes were visually inspected.One of the returned syringes was found damaged/scratched on the top the barrel at 5 unit marking.No damages were found on the second returned syringe.A review of the device history record was completed for batch# 1263405.All inspections and challenges were performed per the applicable operations qc specifications.Based on the samples received, embecta was able to confirm the customer¿s indicated failure on one of the returned syringes.Root cause: from the print carousel, the barrels transition to the transfer dial and onto individual guides that carry the barrels through an oven to cure the ink to the barrel.The printed barrels were getting damaged at the transfer dial to the oven guides.H3 other text : see h10.
|
| |
|
Search Alerts/Recalls
|