MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ PCA ADMINISTRATION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ PCA ADMINISTRATION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 30873
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd alaris¿ pca administration set experienced device damage while considered operable, and leakage.The following information was provided by the initial reporter: bd pca administration set was found by the bedside rn to be cracked at the port that was connected to a patient's tpn and intralipid line.There were no events that led up to the crack.No harm was reported to the patient.The line was replaced with no further problems.
 
Additional Manufacturer Narrative
Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
Investigation summary: one sample was submitted for quality investigation.The customer complaint of component damage - leak was verified by visual inspection.Investigation of the sample revealed a large crack at the female luer adapter connector.A device history record review could not be performed because the lot number is unknown.The root cause for the crack seen in the luer adapter could not be fully determined based on the description of event by the customer.The probable root cause for the damage is that a large amount of force was used to connect luer adapter to its corresponding connector causing it to crack along the length of the adapter.Additionally, if luer connector was cleaned with an antiseptic prior to use, the material that luer is made from can weaken the strength of the luer body, making it easier to crack.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd alaris¿ pca administration set experienced device damage while considered operable, and leakage.The following information was provided by the initial reporter: bd pca administration set was found by the bedside rn to be cracked at the port that was connected to a patient's tpn and intralipid line.There were no events that led up to the crack.No harm was reported to the patient.The line was replaced with no further problems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ PCA ADMINISTRATION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15936694
Report Number9616066-2022-01888
Device Sequence Number15051590
Product Code FPA
UDI-Device Identifier50885403273948
UDI-Public50885403273948
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K811885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number30873
Device Catalogue Number30873
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date12/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-