• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHA
Medical Device Problem Codes Peeled/Delaminated (1454); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation of adhesive peeling of the curved part was confirmed.There were no abnormalities with the appearance and operation of the subject device.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
An olympus employee reported on behalf of the customer, the adhesive on cysto-nephro videoscope was peeling at the tip bending part and the internal wire was exposed during maintenance.Similar issues occur frequently at the facility and recently cyf-va2 was repaired.The facility has 2x cyf-vha and 2x cyf-va, but the same problem is occurring with each.The devices are primarily cleaned and disinfected with endoscope reprocessor (oer-4).At the end of the day go through an ethylene oxide gas sterilization.There was no report of patient harm associated with this event.Event 1 of 4 this report is being submitted for the issue with the 1:2 cysto-nephro videoscope (cyf-vha), under the medwatch with patient identifier (b)(6).The issue with 2:2 cysto-nephro videoscope (cyf-vha) device is being reported on the medwatch with patient identifier (b)(6).The issue with 1:2 visera cystovideoscope (cyf-va) device is being reported on the medwatch with patient identifier (b)(6).The issue with 2:2 visera cystovideoscope (cyf-va) device is being reported on the medwatch with patient identifier (b)(6).
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to the bending section cover has a brush and the cross section is not flat, so we presume that the brush resulted in the lack of bending section cover.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15936867
Report Number3002808148-2022-05000
Device Sequence Number20386055
Product Code FAJ
UDI-Device Identifier04953170310508
UDI-Public04953170310508
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OER-4.
Patient SexUnknown
-
-