| Model Number |
CYF-VHA |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
An olympus employee reported on behalf of the customer, the adhesive on cysto-nephro videoscope was peeling at the tip bending part and the internal wire was exposed during maintenance.Similar issues occur frequently at the facility and recently cyf-va2 was repaired.The facility has 2x cyf-vha and 2x cyf-va, but the same problem is occurring with each.The devices are primarily cleaned and disinfected with endoscope reprocessor (oer-4).At the end of the day go through an ethylene oxide gas sterilization.There was no report of patient harm associated with this event.Event 2 of 4 this report is being submitted for the issue with the 2:2 cysto-nephro videoscope (cyf-vha), under the medwatch with patient identifier (b)(6).The issue with 1:2 cysto-nephro videoscope (cyf-vha) device is being reported on the medwatch with patient identifier (b)(6).The issue with 1:2 visera cystovideoscope (cyf-va) device is being reported on the medwatch with patient identifier (b)(6).The issue with 2:2 visera cystovideoscope (cyf-va) device is being reported on the medwatch with patient identifier (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to scratches of the distal sheath - the the cause of the scratches could not be further presumed.This event was concluded to have not been due to misuse by the user.As the actual product was not returned to the manufacturer, the actual product could not be confirmed/inspected.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|