ZIMMER SURGICAL SA UNIVERSAL MODULAR ELECTRIC/BATTERY DOUBLE TRIGGER HANDPIECE Back to Search Results
Model Number N/A
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: universal battery for aseptic transfer kit (item#: 89-8510-440-20, lot#: 5010595).Foreign source: germany.Multiple mdr reports were filed for this event, please see associated reports: 0008031000 - 2022 - 00057.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that a defect was discovered when the machine was checked at the beginning of the operation.Despite the replacement of the battery and the battery housing, the machine did not work and since no more machines were available that day, the operation had to be cancelled.The patient was already under anesthesia and had to be discharged.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Upon receipt, it was confirmed that the motor was noisy, and the controller board wires were unsoldered/damaged.Trigger card calibration fail - raw material issue.The reported event could not be reproduced.As of repair, motor was replaced with motor with screws, potted wired controller with screws and seals, motor heat shrink kit.Device is used for treatment.Complaint /repair history record review was performed, and no issue was discovered that could explain the defect reported.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL MODULAR ELECTRIC/BATTERY DOUBLE TRIGGER HANDPIECE
Common Device Name
HANDPIECE
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ   CH-1228
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key15938134
Report Number0008031000-2022-00056
Device Sequence Number6394377
Product Code HAB
UDI-Device Identifier00889024354593
UDI-Public(01)00889024354593(11)151009(10)5007183
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number89-8507-400-00
Device Lot Number5007183
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer01/05/2023
03/01/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date01/30/2023
03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
-
-