• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT3 III; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT3 III; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE Back to Search Results
Model Number FT3 G3
Medical Device Problem Codes Incorrect Measurement (1383); High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys ft3 iii and elecsys ft3 iii ver.2 results for 1 patient sample on a cobas e801 module compared to a wako accuraseed analyzer and an abbott architect analyzer.Refer to the attachment to the medwatch for all patient data.No questionable results were reported outside of the laboratory.The customer reagent lots and expiration dates were requested but not provided.The investigational ft3iii lot is 583301 and the expiration date is 28-feb-2023.The investigational ft3iii ver.2 lot is 611920 and the expiration date is 31-may-2023.The customer e801 analyzer serial number is (b)(4).The investigational e801 analyzer serial number is (b)(4).This medwatch will cover ft3iii.Refer to medwatch with patient identifier (b)(6) for information on the ft3iii ver.2 results.
 
Additional Manufacturer Narrative
The patient's sample was provided for investigation.The investigation found that there is no interference with the streptavidin component of the reagent in the sample.It was also determined that there is no interfering factor against the ft3 antibody idiotype.The influence of prewash use was also not observed.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT3 III
Common Device Name
RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15939197
Report Number1823260-2022-03937
Device Sequence Number6442594
Product Code CDP
UDI-Device Identifier04015630939718
UDI-Public04015630939718
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963127
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberFT3 G3
Device Catalogue Number07027362190
Device Lot Number583301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer01/23/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
-
-