SYNTHES GMBH FEMORAL NECK SYSTEM INSERTION HANDLE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 03.168.008
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter name and address: (b)(6).Device evaluated by mfr, device manufacture date: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent surgery with the products in question.After the surgeon inserted the guidewire and finished reaming implant insertion, he noticed that the black screw on the inserter was loose while inserting the distal screw.He tried to re-tighten it, but it had a strange bite and could not be tightened any further.He removed the screw, disconnected the insertion handle connecting the implant, and was instructed to completely remove the black screw and gently turn it in.It was stuck for a while, but after several attempts, he was finally able to turn it in correctly.The procedure was then completed successfully with the reinsertion of the implant and the screws.There was a delay of 30 minutes.No further information is available.This report involves one femoral neck system insertion handle.This is report 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL NECK SYSTEM INSERTION HANDLE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15939284
Report Number8030965-2022-10797
Device Sequence Number17694785
Product Code KTT
UDI-Device Identifier07611819657342
UDI-Public07611819657342
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172872
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.168.008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2022
Initial Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INS F/INSERTION HANDLE; UNK - GUIDE/COMPRESSION/K-WIRES; UNK - SCREWS: TRAUMA
Patient SexUnknown
-
-