• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24690
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The patient was presented with intra-stent restenosis in c1 segment and underwent arterial balloon dilation and a stent implantation.The 95% stenosed target lesion was located in the right internal carotid artery.An 8f boston scientific (bsc) guide catheter was engaged and a filterwire ez guidewire was crossed the lesion.A 4.0mmx30mmx135cm (4f) sterling balloon catheter was advanced for pre-dilatation.During the procedure, the balloon was dilated at 5 atmospheres (atm); however, the physician noticed that the balloon cannot be dilated any further and the pressure was decreasing.The device was removed from the patient by normal operation and found that the balloon has ruptured.Pre-dilation was completed with another of same device and the stent implantation was completed normally.There were no complications and injuries reported and the patient was fine and stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The patient was presented with intra-stent restenosis in c1 segment and underwent arterial balloon dilation and a stent implantation.The 95% stenosed target lesion was located in the right internal carotid artery.An 8f boston scientific (bsc) guide catheter was engaged and a filterwire ez guidewire was crossed the lesion.A 4.0mmx30mmx135cm (4f) sterling balloon catheter was advanced for pre-dilatation.During the procedure, the balloon was dilated at 5 atmospheres (atm); however, the physician noticed that the balloon cannot be dilated any further and the pressure was decreasing.The device was removed from the patient by normal operation and found that the balloon has ruptured.Pre-dilation was completed with another of same device and the stent implantation was completed normally.There were no complications and injuries reported and the patient was fine and stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole in the balloon 24mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a pinhole in the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15939671
Report Number2124215-2022-51298
Device Sequence Number20419759
Product Code DQY
UDI-Device Identifier08714729756972
UDI-Public08714729756972
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141150
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2025
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0029858224
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer03/07/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight76 KG
-
-