• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. LEMAITRE VALVULOTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. LEMAITRE VALVULOTOME Back to Search Results
Model Number 1009-00
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation; therefore, a root cause could not be determined.It was reported that the device damage was observed before using the valvulotome for a second insertion, no damage was noted prior to use by the surgeon.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.Additionally, we have not received any reports of similar issues occurring for this lot.
 
Event or Problem Description
Patient condition prior to surgery was listed as "critical limb ischemia of right lower extremity with gangrene".The case started with an angiogram of the right lower extremity and then progressed to a femoral-tibial bypass grafting using vein harvested from the ipsilateral side.Dr.Gilbertson used the valvulotome once to cut valves within the harvested vein graft.Later in the case the valvulotome was needed again but before attempting to insert it into the vein graft surgical staff discovered that the blade had detached from the tip of the device and was bent slightly.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEMAITRE VALVULOTOME
Common Device Name
VALVULOTOME
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
jeffrey oddy
63 second ave
burlington, MA 01803
7814251682
MDR Report Key15940257
Report Number1220948-2022-00131
Device Sequence Number6404442
Product Code MGZ
UDI-Device Identifier00840663106653
UDI-Public(01)00840663106653
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K140042
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1009-00
Device Catalogue Number1009-00
Device Lot NumberELVH2116V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Initial Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-