In doing count and told that lot c1958996 (that was billed yesterday ) had a problem.Pc = they went down the scope with the stent.When they started to deploy it wouldn't deploy.Stent appears jammed.Product returning for analysis.As per cc form: "the stent was deployed over the wire down the scope but when the nurse started to deploy it it would not go past a certain point regardless if it was retracted or not." a section of the device did not remain inside the patient¿s body.The patient did require any additional procedures due to this occurrence.They had to use an additional stent to complete the case according to the initial reporter, the patient did not experience any adverse effects due to this occurrence at what stage of the procedure did the complaint occur? when unpacking or preparing the evolution, while inserting the evolution in the patient, during stent placement, while removing the introducer, or during stent repositioning/removal during placement.What endoscope type and channel size was used? aquilant duodenal scope serial number (b)(4).What was the position of the elevator? n/a, open, closed n/a.Details of the wire guide used (diameter, type, make)? awg2 0.35.Was the zip port facing upwards and slightly curved when backloading the wire guide? n/a, yes, no.N/a.Did any part of the stent contact the patient¿s anatomy when the complaint occurred? n/a, yes, no.Yes.Please advise the anatomical location of the intended target site.Cbd.How long was the stent in the patient by the time this complaint occurred? a few minutes.For devices where the ifu states for longer term patency has not been established, was periodic evaluation completed? n/a, yes, no n/a.If yes, how often was this completed? did the patient require any additional procedures as a result of this event? n/a, yes, no, yes.What intervention (if any) was required? a further stent was required was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day / same day.Were any other defects (other than the complaint issue) observed on the device prior to return (e.G.Kink)? n/a, yes, no.No 15.If yes, please specify what was observed and where on the device it was observed.Stricture information: 1.What was the length and diameter of the stricture? not known where was the stricture located in the body? cbd.Was there resistance felt passing wire guide through stricture? n/a, yes, no no.Was there resistance felt passing the evolution through stricture? n/a, yes, no no.Was the stricture dilated before stent placement? n/a yes, no.No.Questions related to during insertion into patient was the product inspected for kinks or damage before use? n/a, yes, no yes.Was resistance felt during insertion into patient? n/a, yes, no no.If yes, at what point? questions related to during stent placement did the product fail during stent deployment or recapture? n/a, deployment, recapture, other no.If other, please specify was the directional button pressed during use? n/a, yes, no yes.Was any part of the stent observed in contact with the patient¿s anatomy at the time of failure? n/a, yes, no/ yes.Was the yellow marker kept in view during deployment? n/a, yes, no yes.6.Are images of the device or procedure available? n/a, yes, no] device will be returned questions related to during introducer withdrawal are images of the device or procedure available? n/a, yes, no device will be returned.Was final stent placement confirmed using endoscopy / fluoroscopy? n/a, yes, no.No.If yes, what was used? did the stent open sufficiently to allow withdrawal of introducer safely? n/a, yes, no no.Was the safety wire fully removed before removing the delivery system? n/a, yes, no no.6.Did any part of the product snag/get caught with the stent when removing the delivery system? n/a, yes, no yes all of it as they could not deploy it.Questions related to during stent repositioning/removal (for evo-fc & evo-pc devices) what instrument was used for stent repositioning / removal? forceps, snare, other n/a.If other, please specify.Was resistance encountered during advancement and/or deployment? n/a, yes, no.Yes.If yes, please when this was felt? advancement or deployment.Deployment how did the physician deal with this resistance? removed stent and started procedure with a new one was the lasso (suture) loop used during repositioning no.
|