| Model Number |
3300 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; data review revealed that no error/fault codes had been recorded.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive during pacing threshold testing.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.
|
| |
|
Event or Problem Description
|
|
It was reported that the programmer's screen stopped responding when measuring the threshold of a cardiac resynchronization therapy defibrillator (crt-d) patient.The programmer will be returned for repair and analysis.No adverse patient effect was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; data review revealed that no error/fault codes had been recorded.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive during pacing threshold testing.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.We have received reports of similar events where the latitude programmer screen became unresponsive to touch inputs during use.In most cases, the programmer required a system reboot to complete testing.Representatives from our post market quality assurance, manufacturing, and design assurance groups are actively investigating this unresponsive touchscreen behavior with the latitude programmer to determine the root cause.Based on our initial investigation, this behavior is most likely design-related.
|
| |
|
Event or Problem Description
|
|
It was reported that the programmer's screen stopped responding when measuring the threshold of a cardiac resynchronization therapy defibrillator (crt-d) patient.The programmer will be returned for repair and analysis.No adverse patient effect was reported.
|
| |
|