MERIT MEDICAL SYSTEMS MEXICO MERIT S-MAK¿. COAXIAL MINI ACCESS KIT; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Catalog Number S-MAK401N/C
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is not expected to return for evaluation.A lot number has been provided and a review of the manufacturing history record is in progress.A follow-up report will be submitted once the evaluation is complete.
 
Event or Problem Description
The account alleges that during a av-fistula declot procedure, the access catheter detached from the hub inside the patient's arm.The account alleges that the catheter may have been caught on something within the fistula the way it stretched before it broke.The patient was transferred to another facility for surgical removal of the foreign body.The patient tolerated the procedure well.No additional patient consequences to report.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT S-MAK¿. COAXIAL MINI ACCESS KIT
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego, CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego, CA 92154
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
south jordan, UT 84095
MDR Report Key15941614
Report Number3011642792-2022-00111
Device Sequence Number17657229
Product Code DRE
UDI-Device Identifier00884450012635
UDI-Public884450012635
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K031691
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2025
Device Catalogue NumberS-MAK401N/C
Device Lot NumberI2496601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTERVENTIONAL GUIDEWIRE
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient SexUnknown
-
-