| Model Number |
30914 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd alaris¿ extension sets experienced leakage.The following information was provided by the initial reporter: possible leakage between the valve and tubing.I am not sure if its because we are using two different products.The valve is an icu product connected to the bd tubing.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 22049017, medical device expiration date: 04-01-2025, device manufacture date: 2022-04-01.Medical device lot #: 22019363, medical device expiration date: 2025-01-13, device manufacture date: 2022-01-12.Medical device lot #: 22069500, medical device expiration date: 28-jun-2025, device manufacture date: 22-jun-2021.Medical device lot #: 22049066, medical device expiration date: 05-apr-2025, device manufacture date: 05-apr-2022.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 01-feb-2023.H6: investigation summary: 12 used 30914 extension sets were received and tested by our quality team.The sets were individually tested and 8 of the 12 were occluded when infusion of saline was attempted.This verified the customer's complaint.A digital microscope was then employed to determine the root cause which was an excess application of solvent at the female luer during the production of this set.A device history record review was performed and there were no relevant production issues or quality notifications.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd alaris¿ extension sets experienced leakage.The following information was provided by the initial reporter: possible leakage between the valve and tubing.I am not sure if its because we are using two different products.The valve is an icu product connected to the bd tubing.
|
| |
|