| Model Number |
72290037 |
| Medical Device Problem Code |
Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference:(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy procedure, the ablate function of the werewolf wand was activated although the button was not pressed.The procedure was successfully completed with no delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3, h6: the reported device was received for evaluation.A visual inspection found that the device was returned outside of packaging.No packaging was returned.A visual inspection found no damage to the device.A functional evaluation found the wand was activated previously and experienced a multiple button pressed notification.A functional evaluation found the wand did not activate without the buttons being pressed, but when pressed the buttons did not always activate the wand.This failure has been determined to be related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include: 1) the wand´s connector or cable got damaged.2) saline/humidity entered the interior of the handle shorting the connection of the buttons.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|