• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG UH400 AUTOCON® III 400; ELECTROSURGICAL UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG UH400 AUTOCON® III 400; ELECTROSURGICAL UNIT Back to Search Results
Model Number UH400U
Medical Device Problem Code Energy Spectrum Incorrect (1210)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported, that during two turp cases the surgeon was unable to stop arterial bleeding, and the fulguration was extremely poor.The surgery time of both cases was prolonged.The patient treated during the first case needed to be reoperated, as he was still bleeding, and needed to stay in the hospital a few hours longer.Both patients went home the same day.This report represents the first of the two cases.
 
Additional Manufacturer Narrative
The device was returned and will be forwarded to the manufacturing site in germany for evaluation.This event is filed under internal complaint (b)(4).
 
Additional Manufacturer Narrative
Per manufacturer evaluation findings: based on our initial investigation: by reviewing the similar complaint there is no indication for an increase/ trend of the reported issue.Customer complaint detected: - no.- no relevant (error) entries in the log file of the autocon.Defects found: ---chassis slightly dented in the area of the foot switch sockets.Conclusion and root cause determination: ---no error could be detected.The most probable cause is a defect in the periphery used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UH400 AUTOCON® III 400
Common Device Name
ELECTROSURGICAL UNIT
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM  
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key15942356
Report Number9610617-2022-00328
Device Sequence Number17694678
Product Code GEI
UDI-Device Identifier04048551378809
UDI-Public04048551378809
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171717
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUH400U
Device Catalogue NumberUH400U
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer03/16/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date03/24/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient SexMale
-
-