• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. ATLAS TRANSFER CARRIAGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. ATLAS TRANSFER CARRIAGE Back to Search Results
Model Number AX0660230830000013
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite and found the front latches on the transfer carriage to be bent.The technician stated that the front latches were bent due to user facility personnel not properly aligning the transfer carriage in the sterilizer.As the front latches were bent, the transfer carriage was unable to properly engage to the sterilizer subsequently causing the reported event to occur.Steris discussed the reported event with user facility personnel and the importance of properly aligning the transfer carriage within the sterilizer.The transfer carriage is not under a steris service contract and is serviced and maintained by the user facility.The transfer carriage has been removed from service and is pending repairs.A follow-up report will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that their atlas transfer carriage did not stay engaged within the sterilizer and subsequently contacted an employee.The employee obtained a burn and applied a band-aid.
 
Additional Manufacturer Narrative
In lieu of repairs, the user facility has purchased a new transfer carriage.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS TRANSFER CARRIAGE
Common Device Name
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15943369
Report Number3005899764-2022-00060
Device Sequence Number6219812
Product Code FLE
UDI-Device Identifier00724995150631
UDI-Public00724995150631
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAX0660230830000013
Device Catalogue NumberAX0660230830000013
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer11/10/2022
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-