• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-505
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cartridge was discarded and not received for evaluation.A photo was provided for evaluation which confirmed the venous luer was broken.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet".
 
Event or Problem Description
A report was received on (b)(6) 2022 from the critical care nurse (ccn) of a patient of unspecified pathology, who stated blood was observed leaking from the cartridge during a continuous renal replacement therapy (crrt) on (b)(6) 2022.Additional information was received on 29 nov 2022 from the ccn who stated the patient did not experience any symptoms or require medical intervention.Following the event the patient continues to treat using the nxstage system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Common Device Name
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key15944104
Report Number3003464075-2022-00082
Device Sequence Number12316994
Product Code KDI
UDI-Device IdentifierM535CAR5050
UDI-Public+M535CAR5050/$$072320178033V
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K133547
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2023
Device Model NumberCAR-505
Device Catalogue NumberCAR-505
Device Lot Number20178033
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-