BOSTON SCIENTIFIC CORPORATION CRE PULMONARY; BRONCHOSCOPE ACCESSORY Back to Search Results
Model Number M00550330
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used in the windpipe during a dilatation of windpipe procedure performed on (b)(6) 2022.During the procedure, it was noticed that the balloon could not continue to rise as saline was faintly outflowing from the side.The device was then withdrawn from the patient and a hole was noted.The procedure was completed successfully with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Device code a041001 captures the reportable event of balloon pinhole.Investigation result the returned cre pulmonary dilatation balloon was analyzed and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed, and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the body of the balloon.Microscopic inspection found the balloon had a pinhole located approximately 37 mm from the tip.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon pinhole was confirmed.It is possible that factors encountered during the procedure, such as the interaction with scope and other sharp devices during the procedure could create friction on the balloon and cause a pinhole.Therefore, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used in the windpipe during a dilatation of windpipe procedure performed on (b)(6) 2022.During the procedure, it was noticed that the balloon could not continue to rise as saline was faintly outflowing from the side.The device was then withdrawn from the patient and a hole was noted.The procedure was completed successfully with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PULMONARY
Common Device Name
BRONCHOSCOPE ACCESSORY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15944354
Report Number3005099803-2022-07221
Device Sequence Number6219835
Product Code KTI
UDI-Device Identifier08714729456216
UDI-Public08714729456216
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K023337
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2024
Device Model NumberM00550330
Device Catalogue Number5033
Device Lot Number0027861596
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight65 KG
-
-