• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Halo (2227)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Doctor reported that after implanting an intraocular lens (iol) into the patient¿s left eye they experienced halos and capsular phimosis.The doctor noticed the capsular phimosis when they were explanting the iol.The iol was replaced with a different model and diopter diu150 15.0.The explanted iol will not be returned for evaluation.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.A search of complaints revealed 1 other complaint has been received for this production order number and the issue could not be confirmed as related to manufacturing; therefore, no escalations are required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15944555
Report Number3012236936-2022-02833
Device Sequence Number6210305
Product Code MFK
UDI-Device Identifier05050474751187
UDI-Public(01)05050474751187(17)240302
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/02/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Initial Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-