| Model Number |
SOFTECI+23.0 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Lenstec received an email stating " once lens was implanted, haptics were amputated, removed, implanted same diopter hd lens.No patient injury but suture enlarged.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the review of the batch documentation conducted, lenstec can confirm that all procedures in the manufacturing and packaging of the devices were conducted correctly.Additionally, we have not received any other complaints from these batches.Furthermore, after device evaluation it was determined that the damage to the haptics suggests that the incorrect technique was used to load and eject the lens, indicating the manufacturers instructions for use were not followed.
|
| |
|
Search Alerts/Recalls
|