| Model Number |
68-150 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformance's noted.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right-side "firm" and capsular contracture, baker grade iv.Device remains implanted.
|
| |
|
Search Alerts/Recalls
|