BOSTON SCIENTIFIC CORPORATION CRE PULMONARY; BRONCHOSCOPE ACCESSORY Back to Search Results
Model Number M00550340
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used in the trachea during a balloon trachea dilatation procedure performed on (b)(6) 2022.During the procedure, it was noticed that the pressure did not continue to rise after being connected to the pressure gun for inflation.Reportedly, saline in the balloon leaked out from the side.The device was retrieved from the patient and holes were noted on the balloon.The procedure was completed with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used in the trachea during a balloon trachea dilatation procedure performed on (b)(6) 2022.During the procedure, it was noticed that the pressure did not continue to rise after being connected to the pressure gun for inflation.Reportedly, saline in the balloon leaked out from the side.The device was retrieved from the patient and holes were noted on the balloon.The procedure was completed with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: device code a041001 captures the reportable event of balloon pinhole.Block h10: investigation result: the returned cre pulmonary dilatation balloon was analyzed and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the balloon body.Microscopic inspection found the balloon had a pinhole located approximately 32mm from the tip.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon pinhole was confirmed.It is possible that factors encountered during the procedure, such as the interaction with scope and other sharp devices during the procedure could create friction on the balloon and cause a pinhole.Therefore, the most probable root cause is adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PULMONARY
Common Device Name
BRONCHOSCOPE ACCESSORY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15944960
Report Number3005099803-2022-07283
Device Sequence Number6347231
Product Code KTI
UDI-Device Identifier08714729456223
UDI-Public08714729456223
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K023337
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2025
Device Model NumberM00550340
Device Catalogue Number5034
Device Lot Number0028918301
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
Patient Weight62 KG
-
-