Model Number DIU225
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when inserting an intraocular lens (iol) into the patient¿s left eye, the lens did not deploy because the device tip was bent.The tip possibly touched the patient.Reportedly, this is not a stuck in cartridge issue.The issue is related to the bent tip.It came out of the package that way.The issue is not related to use error.The procedure was completed successfully with a backup iol.There were no complications such as incision enlargement, vitrectomy, or sutures.No further information was provided.
 
Additional Manufacturer Narrative
Weight and ethnicity: information unknown/not provided.Implant date, if implanted, give date: not applicable, as the device was not implanted.Explant date, if explanted, give date: not applicable, as the device was not implanted, therefore, not explanted.Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.The search of complaints revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: dec 19, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: the complaint handpiece was received inside of the original folding carton.Visual inspection under magnification revealed that the complaint handpiece was received with the lens stuck inside of the cartridge tip.The cartridge was cracked and damaged around the lens, the cartridge tip was cracked as well.The lens could be observed to be overridden by the plunger rod.The handpiece assembly was inspected and, no issues that could contribute to or cause the observed issue could be identified.A product quality deficiency could not be determined.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15945268
Report Number3012236936-2022-02856
Device Sequence Number6442798
Product Code MJP
UDI-Device Identifier05050474747050
UDI-Public(01)05050474747050(17)240217
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2024
Device Model NumberDIU225
Device Catalogue NumberDIU225U160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer01/03/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1