Model Number BEA28-80/I20-40
Medical Device Problem Codes Material Puncture/Hole (1504); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924); Clinical Endoleak (4843)
Date of Event 11/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft, an afx vela suprarenal proximal extension, and an afx limb stent graft distal extension.Approximately five and a half (5.5) years post initial procedure, the patient presented with a potential type iiib endoleak as detected per ct (computed tomography) scan.The physician elected to treat the patient by implanting two (2) ovation ix distal extenders on (b)(6) 2022.The endoleak was successfully resolved and the patient is reportedly stable.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type iiib endoleak (of the bifurcated stent graft) and additional endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest aneurysm sac growth (of 14.6mm) occurred that was not included in the event as reported.This finding was discovered during an examination of the 64-month post implant ct (computed tomography) scan.It was also noted that there was a rupture pre-index procedure, which is a cautionary product use condition.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported to be in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: h6 health effect - clinical code; remove 1924.H6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft, an afx vela suprarenal proximal extension, and an afx limb stent graft distal extension.Approximately five and a half (5.5) years post initial procedure, the patient presented with a potential type iiib endoleak as detected per ct (computed tomography) scan.The physician elected to treat the patient by implanting two (2) ovation ix distal extenders on (b)(6) 2022.The endoleak was successfully resolved and the patient is reportedly stable.Additional information: clinical assessment identified significant aneurysm enlargement (greater than 5mm) also occurred in addition to a pre-existing aortic rupture (cautionary product use condition).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15945980
Report Number2031527-2022-00275
Device Sequence Number12312218
Product Code MIH
UDI-Device Identifier00818009014412
UDI-Public(01)00818009014412(17)170526
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 11/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/26/2017
Device Model NumberBEA28-80/I20-40
Device Lot Number1499830-031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date02/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/15/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX LIMB STENT GRAFT, SN (B)(6).; AFX VELA SUPRARENAL, SN (B)(6).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-