MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
SHPX260 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) on lot 9610063 was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast surgery with a 260cc mentor memorygel xtra breast implant and experienced capsular contracture (baker grade 3) on an unknown side post-operatively.As a result, the patient underwent explantation on (b)(6) 2022.Two device serial numbers (b)(4), have been documented, however, it is unknown which serial number belongs to the impacted side.If more information becomes available, mentor will submit a supplemental report accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
On december 29, 2022, the mentor failure analysis lab has received the device for evaluation.On january 04, 2023, the evaluation for the device was completed.Device evaluation summary: during the visual evaluation, no apparent damage or visual anomalies were observed on the smooth high profile xtra 260cc returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On january 17, 2023, mentor received additional information regarding the patient, impacted device, and impacted side.The patient's weight has been updated to 164 lbs.It was reported that the impacted side was the patient's right side.The serial number for the right side was determined to be (b)(6).Section d4 has been updated to reflect this information.Manufacturer¿s reference number: (b)(4).
|
| |
|