SMITH & NEPHEW, INC. JOURN ART INS BCS STD 7-8 LT 9; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74023281
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, a patient had an unstable left knee, so a revision surgery was booked on the (b)(6) 2022 to explant one (1) journal art ins bcs std 7-8 lt 9 which had a post broken.A 7/8 13mm journey i was inserted in exchange.No delay reported.Patient was not harmed as a consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, there were no clinical factors found which would have contributed to the event.Based on the information provided, the impact to the patient was the reported fracture and subsequent revision.According to the report there was no delay in the procedure and the patient was not harmed because of this problem.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.According to the material specification, the quality and manufacture of ultra-high-molecular-weight polyethylene (uhmwpe) as ram-extruded rod and compression-molded rod and plate shall be controlled.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that in warnings and precautions that the implant can break or become damaged as a result of strenuous activity or trauma.The patient should be warned that the device does not replace normal healthy bone.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, size selected, surgical technique or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURN ART INS BCS STD 7-8 LT 9
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15946243
Report Number1020279-2022-04912
Device Sequence Number20460641
Product Code JWH
UDI-Device Identifier03596010555342
UDI-Public03596010555342
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2017
Device Model Number74023281
Device Catalogue Number74023281
Device Lot Number07EM01988
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date12/08/2022
Supplement Report FDA Received Date01/11/2023
Date Device Manufactured05/31/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-