| Model Number |
71118230 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery one (1) 11.0mm fixed endcutting rmr head hit a blocking wire causing the shaft of the reamer to spiral split all the way up.No pieces fell into the wound.The shaft broke outside the patient while the reamer head was inside patient so nothing broke into the patients wound.The tossed quickly after the case.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery one (1) 11.0mm fixed endcutting rmr head hit a blocking wire causing the shaft of the reamer to spiral split all the way up.No pieces fell into the wound.The shaft broke outside the patient while the reamer head was inside patient so nothing broke into the patients wound.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|