• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VAMF3636C200TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Relevant device(s) are: product id: v amf3636c200tu, serial/lot #: (b)(4), ubd: 15-apr-2020, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant navion stent graft system was implanted in a patient for the endovascular treatment of an unknown size thoracic aortic aneurysm.A valiant captivia stent graft was implanted approximatley 1 year later.It was confirmed that the most proximal stent graft was the valiant navion vnmf4646c103tu implanted as an extension placed inside the valiant captivia vamf4036c150tu.It was reported approximately 4 years post the index proceure the patent presented emergently to the hospital with unknown symptoms , where a ct was performed and revealed an endoleak  in the proximal descending aorta where 4 endografts were previously placed.The patient was brought to or where a angiogram in or revealed a type iii endoleakcoming from 2 of the most proximal stent grafts , a valiant navion vnmf4646c103tu & a valiant captiva vamf4036c150tu.The physician stated at the time it was either a disjunction and/or stent fracture.The event was treated on the same day with 3 valiant captivia implanted to reline the thoracic aorta.As per the physician, cause of the event cannot be determined. no additional clinical sequalae were reported and the patient is fine corelab review of images from (b)(6) 2021 for the navion device saw no aortic enlargement, endoleak, stent ring enlargement or stent ring migration.Fracture was non-evaluable.Core lab review of ct imaging from (b)(6) 2022 for the navion stent graft identified no endoleak, no stent ring enlargement or no aortic enlargement.The imaging was noted as ne for stent fracture and ne for stent ring migration due to overlaps.A type ib endoleak on a non-navion device at distal end of treated area and endoleak of unknown type on non-navion device in proximal descending thoracic aorta were observed.It was confirmed it is difficult to determine the location and number of additional devices present.3 additional devices were noted, but there are likely more.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15948353
Report Number9612164-2022-04650
Device Sequence Number12596333
Product Code MIH
UDI-Device Identifier00643169346567
UDI-Public00643169346567
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2020
Device Model NumberVAMF3636C200TU
Device Catalogue NumberVAMF3636C200TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Initial Report FDA Received Date12/09/2022
Date Device Manufactured04/16/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-