Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message (error 5) associated to the temperarature problem and stirrer motor during maintenance.In addition, foam was out of the tap overfolow.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: no service activity has been scheduled yet for this device.However, based on investigation results of previous similar cases, the reported error is related to a too-low power consumption of the pump motor, therefore most likely it did not work correctly.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2013.As the event is isolated and the pumps have never been changed, it cannot be excluded that an electro-mechanical defect has led to the displayed error.Based on the fact that it was noted foaming out of overflow (useful for collecting excess water), it cannot be ruled out that another contributing root cause could be related to excessive water in the tanks reaching internal circuits and damaging the impacted pump motor.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15948384
MDR Text Key306898728
Report Number9611109-2022-00643
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/11/2022
Initial Date FDA Received12/09/2022
Supplement Dates Manufacturer Received03/31/2023
Supplement Dates FDA Received04/28/2023
Date Device Manufactured07/23/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-