MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR4054
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 05/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, saline allegedly leaked from the tip of the balloon.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure, saline allegedly leaked from the tip of the balloon.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one dorado pta dilatation catheter with an unknown inflation device and one syringe were returned for evaluation.On visual evaluation of the device, no other specific anomalies were noted.On functional evaluation, the device was inflated with the returned inflation device at 10 atm and maintained pressure; on further inflation up to 20 atm, leaks were noted at the catheter bond joint.Microscopic observation was performed and a break was noted at the proximal glue joint.Four images were reviewed.The four images from an angioplasty procedure in the forearm are reviewed.They demonstrate inflation of an angioplasty balloon in the distal forearm with extravasation of contrast material from the balloon, suggestive of balloon rupture.As per the reported event, the angioplasty balloon leaked contrast during inflation.This is confirmed with the provided images.Therefore, the investigation for the reported leak is confirmed as water was leaking from the bond joint during the evaluation of the returned device.The reported event can be also confirmed and observed on the provided images for review.The investigation is also confirmed for the identified bond joint break as a break at the proximal glue joint was able to be observed under microscopic observation.A definitive root cause for the reported leak and identified bond joint break could not be determined based upon the provided information.H10: d4 (expiry date: 05/2024), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15948605
Report Number2020394-2022-00973
Device Sequence Number6371601
Product Code LIT
UDI-Device Identifier00801741058486
UDI-Public(01)00801741058486
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR4054
Device Catalogue NumberDR4054
Device Lot Number93SF0092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date12/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-