Model Number TFAT30
Medical Device Problem Codes Break (1069); Malposition of Device (2616)
Health Effect - Clinical Codes Blurred Vision (2137); Insufficient Information (4580)
Date of Event 10/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A practice manager reported that after intraocular lens (iol) implantation, the lens broken at haptic optic junction.The lens was explanted and replaced with same model and diopter within 2 weeks after implantation.Additional information has been requested.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received stating that the patient presented for 1 day post operation visit, it was noted that lens was not at desired axis.It was decided to rotate the lens.During the attempt to rotate the patient's original iol, physician noted that the iol haptic optic junction was broken.For this reason, a new toric iol has been placed with same model and diopter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15948789
Report Number1119421-2022-02616
Device Sequence Number7302753
Product Code MFK
UDI-Device Identifier00380652418676
UDI-Public00380652418676
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/01/2025
Device Model NumberTFAT30
Device Catalogue NumberTFAT30.185
Device Lot Number15050145
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date01/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HICSOAP; HOLLADAY 1; IOL WEBBASED CALCULATORACRYSOF; PROVISC OVD; UNSPECIFIED- FORCEP; WANG-KOCH ADJUSTMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
-
-