SMITH & NEPHEW, INC. GII C/R ART INS SZ 3-4 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420490
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a tka, one (1) genesis ii cruciate retaining articular insert size 3-4 9mm broke during cementing/setting.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
 
Event or Problem Description
It was reported that during a tka, one (1) genesis ii cruciate retaining articular insert size 3-4 9mm broke during cementing/setting.All pieces were removed with pick ups and nothing was left in the patient.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
B5: describe event or problem, d4: lot number and h6: health effect - impact code.Section h3, h6: the device was not returned for evaluation.However, the photograph was reviewed, and revealed that the device is broken.The clinical/medical investigation concluded that, this case reports a genesis ii cruciate retaining articular insert size 3-4 9mm broke during cementing/setting.Per the report, all the broken pieces were removed with pickups and nothing was left in the patient.No other clinical information was provided.According to the report, the procedure was completed after a non-significant delay, with a smith+nephew back-up device.It was reported the patient was not injured as consequence of this problem and no further harm is anticipated.Therefore, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that the implant can break or become damaged as a result of strenuous activity or trauma.The patient should be warned that the device does not replace normal healthy bone, this has been identified as warnings and precautions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with the material specification it shall control the quality and manufacture of ultra-high-molecular-weight polyethylene (uhmwpe) as ram-extruded gur 1050 rod and compression-molded gur 1020 rod and plate.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include surgical technique or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.B1: adverse event and/or product problem, b2: outcomes attributed to adverse event, h1: type of reportable event , h5: labeled for single use and h8: usage of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII C/R ART INS SZ 3-4 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15948836
Report Number1020279-2022-04927
Device Sequence Number7334822
Product Code JWH
UDI-Device Identifier03596010207890
UDI-Public03596010207890
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420490
Device Catalogue Number71420490
Device Lot Number10JM16551
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-