| Model Number |
M3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Limb Fracture (4518)
|
| Date of Event |
11/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reports provided indicate when the end-user was in the shower, they reportedly lost their balance and fell forward out of the seating to the floor of the shower.This fall reportedly resulted with the end-user having sustained a fracture to their left tibia.Reportedly the end-user utilizes the device in this manner on a routine basis while keeping the device covered to protect from moisture.Service provider stated having inspected the device and found it to be operating normally with no signs of a malfunction having occurred.The service provider hypothesized the end-user inadvertently activated the seat functions into an anterior tilt, and this caused the end-user to lose positioning.As no claims or allegations were received of a product malfunction having occurred, and the device remaining fully operational after the event; permobil in unable to reach a determination as to cause of adverse event without speculation.The dhr was reviewed, and device was found to have met specification prior to distribution.
|
| |
|
Event or Problem Description
|
|
Report received claiming the end-user was using their power wheelchair inside a roll in shower.Report claims during the showering activities, at some point the end-user lost their positioning and fell out of the seating to the shower floor where they reportedly sustained injuries requiring medical intervention to address.
|
| |
|