• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 60CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 60CM; SCS EXTENSION Back to Search Results
Model Number 3346
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 10/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 1627487-2022-06694.It was reported the patient is not receiving effective therapy due to high impedances.X-rays indicated there was a fracture in the splitter.Surgical intervention is planned to address the issue.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Additional information was received wherein the extensions were explanted and replaced and effective therapy was established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 60CM
Common Device Name
SCS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15949028
Report Number1627487-2022-06693
Device Sequence Number17977172
Product Code LGW
UDI-Device Identifier05414734402309
UDI-Public05414734402309
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2016
Device Model Number3346
Device Catalogue Number3346
Device Lot Number4413940
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer02/05/2023
04/17/2023
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date02/05/2023
04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS IPG; SCS LEADX4
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight98 KG
-
-