• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS SODIUM ELECTRODE; ELECTRODE, ION SPECIFIC, SODIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS SODIUM ELECTRODE; ELECTRODE, ION SPECIFIC, SODIUM Back to Search Results
Catalog Number 21029371001
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer (fse) found a faulty electrode and replaced all electrodes, replaced both probes, and replaced the tubing.Calibration, qc, patient correlations, and an ise performance check were all performed successfully.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
 
Event or Problem Description
There was an allegation of questionable gen.2 ise indirect for na results for 1 patient sample on a cobas integra 400 plus module.The initial na result was 129.9 mmol/l.The result was reported outside of the laboratory.The customer was having issues with out-of-range qc and repeated the sample on another 400-plus analyzer.The repeat result on another analyzer was 136.2 mmol/l.The repeat result was deemed correct.The integra 400 plus analyzer serial number is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SODIUM ELECTRODE
Common Device Name
ELECTRODE, ION SPECIFIC, SODIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15949044
Report Number1823260-2022-03946
Device Sequence Number7302761
Product Code JGS
UDI-Device Identifier04015630032457
UDI-Public04015630032457
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K963627
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21029371001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Initial Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-