U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRONOVA SUTURE 24"(60CM) 7-0 BLU; SUTURE, NON ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRONOVA SUTURE 24"(60CM) 7-0 BLU; SUTURE, NON ABSORBABLE, SYNTHETIC Back to Search Results
Model Number M3735
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient underwent an unknown cardiovascular surgery on (b)(6) 2022 and suture was used.During use, when putting the first stitch and pulling the suture, the suture broke easily.Another product from a different box was used.Further details are not provided.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Additional information provided: when putting the first stitch and pulling the suture, the suture broke easily.Another product from a different box was used.When putting the first stitch and pulling the suture, the suture broke easily.After that, same lot product from unopened package from the same box was used but the suture broke easily again.Therefore, another product from a different box was used for the surgery.Lot number : unk: sbbbbb qty of product involved : 1 : 2 qty to be returned : 1 : 4 attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Note: related events reported on mw# 2210968-2022-10149.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Date sent to the fda: 1/20/2023 h6 component code: g07002 no device problem found additional information: d9, h3, h6 h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one opened straight folder packet with eight needle-suture pieces that pertain to product code m3735 was received for evaluation.The product code should be multistrand and each folder must contain four strands double-armed per packet.In order to evaluate the conditions of the returned sample, the suture pieces were observed with the ends cut.Body fluids and crimping marks were observed on the surface of the suture that were possibly caused by a surgical instrument.The condition of the samples received indicates improper handling of the device and no functional test was performed.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that a closed packet that pertains to product code m3735 was received for evaluation.Upon initial inspection, of the sample, no external damages were observed on the packet.Visual inspection of the unopened sample, no defects were found on the packet.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRONOVA SUTURE 24"(60CM) 7-0 BLU
Common Device Name
SUTURE, NON ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15949252
Report Number2210968-2022-10151
Device Sequence Number20667778
Product Code GAW
UDI-Device Identifier10705031045187
UDI-Public10705031045187
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM3735
Device Catalogue NumberM3735
Device Lot NumberSBBBBB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-