• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1; HIP CEMENTLESS STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1; HIP CEMENTLESS STEM Back to Search Results
Model Number 01.12.021
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery performed at about 11 years 2 months from the primary due to loosening of the stem.The surgery was completed successfully by implanting a minimax.
 
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2022.Lot 100759: (b)(4) items manufactured and released on (b)(6) 2010.Expiration date: 2015-03-31.No anomalies found related to the problem.To date, all items of the same lot have been sold without any similar reported event during the period of review.Clinical evaluation performed by medical affairs department: revision surgery 11 years 2 months after index surgery due to loosening of the stem in a female patient.Radiographic image provided show the presence of radiological mobilized stem and possibly some sort of pathological reaction of the bone.The report does not mention infection.The stem appears small for the canal perhaps due to subsidence of the stem secondary to loosening.The acetabular cup was originally implanted in a very vertical position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1
Common Device Name
HIP CEMENTLESS STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15950502
Report Number3005180920-2022-00913
Device Sequence Number15310111
Product Code JDI
UDI-Device Identifier07630030802027
UDI-Public07630030802027
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2015
Device Model Number01.12.021
Device Catalogue Number01.12.021
Device Lot Number100759
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Initial Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/14/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight65 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-