| Model Number |
153625H |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the temperature sensing extension cord has poor temperature display and the temperature display disappears.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be " input female connector wears out".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "do not autoclave or sterilize the monitor cable by ethylene oxide.Wipe the monitor cable with a soft cloth using a solution of mild detergent and warm water or disinfectant, dry thoroughly." correction: g, h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the temperature sensing extension cord had poor temperature display and the temperature display disappeared.Per follow-up received via ibc on 16dec2022, stated that temperature sensing extension cord had poor display.
|
| |
|