• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US HP MBT CEM KEEL PUNCH SZ 4-7; KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US HP MBT CEM KEEL PUNCH SZ 4-7; KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET Back to Search Results
Model Number 9505-02-012
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that yesterday, while covering one of the surgeon's sigma knee replacement cases, his physician assistant asked the rep what it would take to replacement some of the keel punches in their hospital's consignment instrumentation trays.She then showed the rep how one or both of the small metal tines on the connection end of the keel punch they were about to use were missing.She explained that the punches can still be used, even without the missing tines, but they don't fit as securely on the impaction handles without the missing tines.Sales rep explained our company's device reporting system and instrumentation replacement process and then proceeded to inspect all of the keel punches in the hospital's consignment trays upon the completion of our case.A total of five punches were found with broken tines.It is unknown when and where the tines broke off.The broken punch needed in the surgeon's current case was used without incidence, since an immediate replacement was unavailable.The surgical time of the case was not affected.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the device confirmed the reported allegation.The assembly tab was found broken, the fragment was not returned for evaluation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a date code was provided, which indicates that the device was manufactured on 2009.A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.Corrected h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP MBT CEM KEEL PUNCH SZ 4-7
Common Device Name
KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15951475
Report Number1818910-2022-24948
Device Sequence Number6561382
Product Code LRY
UDI-Device Identifier10603295227328
UDI-Public10603295227328
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9505-02-012
Device Catalogue Number950502012
Device Lot NumberJ0309
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer12/28/2022
01/04/2023
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date12/29/2022
01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-