| Model Number |
VICM5_13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted and removed a 13.2mm vicm5_13.2 implantable collamer lens, -11.50 diopter, from the patients right eye (od) due to the big size of the lens.The lens was replaced with a smaller size lens.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h6: health impact- clinical code: 4581 - significant reduction of irido-corneal angles.Corrected data: b5: the reporter indicated the surgeon implanted a 13.2mm vicm5_13.2 implantable collamer lens, -11.50 diopter, into the patients right eye (od) on (b)(6) 2022.The lens was removed on (b)(6) 2022 due to excessive vaulting and significant reduction of irido-corneal angles.The lens was replaced with a smaller size lens and the problem was resolved.The cause of the event was due to the device.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|