Model Number 105-5091-150
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Healthcare professional information not available due to privacy laws.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report regarding echelon catheter perforation (with guidewire/stylet).The patient was undergoing an aneurysm coil embolization.The accessed vessel was the ic with a vessel diameter of 5mm.It was noted the patient's vessel tortuosity was strong.It was reported that tactics were delivered before the ic-an to ic-an coil embolization.For the microcatheter, the echelon 10st (lot: b360807) was selected, preparation was carried out as per the package insert, and a steam shape of 30 seconds was applied to the j-shape of the echelon 10st.After the shape, when an attempt was made to set up a synchro select standard 14 wire, 14 wires were perforated at a point around 1 cm from the end of the extension.The usage of the echelon10 abstained, and there was no backup available, so the mc (sl10) of another company's product was opened and prepared in the same manner.A steam shape was performed on the j shape in the same way as before, and the setup using the 14-wire was completed without any problems.After that, there were no problems and coil embolization was performed successfully.All products were prepared as per the package insert.The product was not used in the patient's body because abnormalities were confirmed during the setup stage.No residual inside the patient's body.It was reported that catheter perforation (with guidewire/stylet) occurred in the distal shaft.There was no resistance when passing the guidewire/stylet.There were no kinks or other damage on the guidewire, coil, or stylet.The device was prepared as indicated in the package insert.The catheter was flushed as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.Ancillary devices include a synchro select standard 14 guidewire.
 
Event or Problem Description
Additional information received clarified that "ic-an" referred to an internal carotid artery aneurysm.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15953785
Report Number2029214-2022-02103
Device Sequence Number12619062
Product Code KRA
UDI-Device Identifier00836462002241
UDI-Public00836462002241
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093750
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2024
Device Model Number105-5091-150
Device Catalogue Number105-5091-150
Device Lot NumberB360807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date12/21/2022
Date Device Manufactured03/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-