Model Number M00326520
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the mechanical thrombectomy procedure, the subject guidewire was broken into 2 fragments.Medical intervention was performed to extract the broken fragment but was unsuccessful and left inside the patient anatomy.The physician replaced it with a new device and continued the procedure.The procedure was completed successfully.No further information available.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes), there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review of available information failed to identify a definitive cause and the device was not returned for analysis, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during the mechanical thrombectomy procedure, the subject guidewire was broken into 2 fragments.Medical intervention was performed to extract the broken fragment but was unsuccessful and left inside the patient anatomy.The physician replaced it with a new device and continued the procedure.The procedure was completed successfully.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRO2-14 STANDARD PRE-SHAPED 300CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15954243
Report Number3012931345-2022-00222
Device Sequence Number6561957
Product Code DQX
UDI-Device Identifier04546540688798
UDI-Public04546540688798
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K053268
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/04/2025
Device Model NumberM00326520
Device Catalogue NumberM00326520
Device Lot Number0000151278
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-