Catalog Number 955468
Device Problems Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Event Description
It was reported that approximately 2-3 hours into treatment with a prismaflex st100, an external fluid leak was observed in the drain bag.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter address: (b)(6).Prismaflex st100 set c has been temporarily approved for use in the us under emergency use authorization (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.The actual sample was not provided for evaluation; however, a picture was provided.The reported condition was verified.The cause is user error.This issue is being further investigated.The customer reported that an external fluid leakage from the drain bag occurred 2-3hours after treatment started.All accessories provided within the set, including drain bag, are single use products.This means that once used, the drain bag must not be reused (i.E.Emptied and reused during the same therapy) and should be discarded.However, the prismaflex control unit operator's manual indicates to ¿change fluid bags when the appropriate caution alarm occurs (pbp bag empty, replacement bag empty, dialysate bag empty or effluent bag full)¿.Furthermore, the prismaflex control unit warns the operator when the effluent bag is full and it is specified in the on-screen instruction delivered by the prismaflex control unit that the drain bag should be disconnected and changed for a new bag.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET (ST)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7 av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15954703
MDR Text Key305792095
Report Number8010182-2022-00387
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123468
UDI-Public(01)07332414123468
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number955468
Device Lot Number22E0076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2022
Initial Date FDA Received12/09/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-