Catalog Number PNML6F088904
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a thrombectomy procedure using a neuron max 6f 088 long sheath (neuron max) and a guidewire.During the procedure, after completing multiple passes using the neuron max, the physician fractured the neuron max at the proximal end near the hub.Therefore, the neuron max was removed.The procedure had ended at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned neuron max confirmed that the catheter was fractured.Based on the reported event, the device was used to complete multiple passes before it fractured.The manipulation of the catheter during subsequent use, may have contributed to the fracture.Further evaluation revealed kinks in the catheter shaft.This damage was incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15954711
Report Number3005168196-2022-00572
Device Sequence Number6503398
Product Code DQY
UDI-Device Identifier00815948022393
UDI-Public815948022393
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/09/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2025
Device Catalogue NumberPNML6F088904
Device Lot NumberH00000532
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Initial Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-