| Model Number |
IMHA34 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the abutment became loose and doctor tightened and sent patient home.Abutment was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One (1) certain ep healing abutment 3.4mm(d) x 3.8mm(p) x 4mm(h) (imha34) was not returned for investigation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item and lot number along with the applicable device history record (dhr), instructions for use (ifu), and risk file.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.No pre-existing condition noted, and no per form was provided.Bone density type is unknown.The reported device was located on an unknown tooth site (universal) and was used for an unknown amount of time.The customer did not provide any pictures or x-rays.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number (1241643).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (1241643) for similar events and no other complaint was identified.January post market trending was reviewed and there were no actionable events or corrective actions for the reported event (loosening) or product (imha34).No actionable items have been triggered that will affect complaint handling on our end for this month.Therefore, based on the available information, device malfunction and the reported event could not be verified.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|