| Model Number |
X SERIES |
| Medical Device Problem Code |
Incorrect Interpretation of Signal (1543)
|
| Health Effect - Clinical Code |
Necrosis (1971)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a 76-year-old female patient, the device issued a "shock advised" prompt for a heart rhythm they believed was non-shockable.Complainant indicated that the patient subsequently expired.However, it was not a result of the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation for evaluation.Instead, the clinical data files were provided for review.Our evaluation of the clinical files determined that cpr was being performed during the rapid shock analysis period.This resulted in a "shock advised" being recommended.Cpr should be stopped and contact with the patient ceased during an analysis period.Our analysis concluded that the device worked to specification.Analysis of reports of this type has not identified an increase in trend.
|
| |
|