Model Number ORT300
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The leak was discovered during a detailed visual inspection with table covers removed, which is performed as part of preventive maintenance.The leak was traced to a union connection to a stainless steel braided ptfe hydraulic hose within the pressurized hydraulic circuit.The service engineer did not observe any functional or performance impact to the table as a result of the leak, which was minor in nature.The hydraulic hose and fitting were replaced and the table preventive maintenance service was completed.
 
Event or Problem Description
During routine preventive maintenance, an imris service engineer observed leakage of hydraulic fluid from a ptfe hose union within the operating room table.No table functional issues were observed and no reports have been received from the customer relating to the fluid leak.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT300 OPERATING ROOM TABLE
Common Device Name
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
paul campbell
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key15955521
Report Number3010326005-2022-00018
Device Sequence Number12402571
Product Code KXJ
UDI-Device Identifier00857534006462
UDI-Public(01)00857534006462
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberORT300
Device Catalogue Number114093-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Initial Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-