| Model Number |
ORT300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The leak was discovered during a detailed visual inspection with table covers removed, which is performed as part of preventive maintenance.The leak was traced to a union connection to a stainless steel braided ptfe hydraulic hose within the pressurized hydraulic circuit.The service engineer did not observe any functional or performance impact to the table as a result of the leak, which was minor in nature.The hydraulic hose and fitting were replaced and the table preventive maintenance service was completed.
|
| |
|
Event or Problem Description
|
|
During routine preventive maintenance, an imris service engineer observed leakage of hydraulic fluid from a ptfe hose union within the operating room table.No table functional issues were observed and no reports have been received from the customer relating to the fluid leak.There was no patient involvement.
|
| |
|