• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA095901A
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2022, the patient underwent a tips reduction procedure using a gore® viabahn® vbx balloon expandable endoprosthesis.The distal end of the vbx device was unsheathed and the balloon was inflated to expand the distal portion inside previously implanted viatorr stent.The physician then deflated the balloon and attempted to pull the balloon back to complete the remaining portion of the partially deployed vbx device.However, it was reported the delivery system catheter became stuck on the guidewire (amplatz wire).The physician disconnected the hub from the catheter trying to remove the delivery system from patient.The physician then pulled the guidewire and delivery system catheter out of the patient.The stent graft dislodged from the delivery system catheter and remained in its intended location partially deployed.The physician then advance a guidewire into the partially deployed vbx device.The procedure was complete by advancing a balloon and expanding the partially deployed vbx device.The device was reportedly able to be fully expanded in its intended location.No patient harm or impact was reported.
 
Additional Manufacturer Narrative
Engineering evaluation: the device was returned with no endoprosthesis or hub assembly present.The balloon catheter was necked and broken consistent with the application of force during reported hub disconnection in the procedure.The balloon cover appeared constrained on the proximal end with corresponding ring impressions while the appearance of the balloon cover on the distal end was consistent with partial deployment as reported.No further remarkable observations were noted with respect to the delivery system.The gore® viabahn® vbx balloon expandable endoprosthesis instructions for use (ifu) states, ¿during positioning, verify that the endoprosthesis is still centered within the marker bands and has not been dislodged.Do not deploy the endoprosthesis unless it is properly centered on the balloon and properly positioned within the target lesion.If the position of the endoprosthesis within the lesion is not optimal, it should be carefully repositioned or removed.¿ the ifu also states: ¿if needed, the delivery catheter and the sheath or guide catheter may be removed together as a unit.Excessive or abrupt force during catheter removal may damage the delivery catheter, or introducer sheath.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
marcos ayala
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15955713
Report Number2017233-2022-03563
Device Sequence Number15104399
Product Code PRL
UDI-Device Identifier00733132637362
UDI-Public00733132637362
Combination Product (Y/N)Y
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2024
Device Model NumberBXA095901A
Device Catalogue NumberBXA095901A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer11/16/2022
11/16/2022
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date01/04/2023
01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-