| Model Number |
FF-410595 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Fax number: (b)(6).A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and exchange of textured device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation- based on the device analysis grid, the assessments of the complaint are: ¿ rupture : observed broken sharp assessed as an unidentified (tear) opening and missing piece of the shell assessed as to inconclusive.¿ anxiety-product/procedure: unable to confirm as it is a medical event not related to the device.Additional observations: ¿ crease flat observed in the device.
|
| |
|